Spanish  |   English
 
   
 
  - Unit that allows to mix, compressed air and oxygen, automatically adjustable between 21% and 100% of oxygen.

MODEL BL-3500 A
This model has an pneumatic alarm, that is also activated when the difference of pressure between both gases is more than 1,5 Kg/Cm2.
When this alarm is triggered, the mixer automatically opens a bypass valve, that allows continue function with the remaining gas.
Adjustable flow with a flowmeter from 2 to 32 LPM

MODEL BL-3500
Idem Model BL-3500 A, without pneumatic alarm and bypass valve.

POWER SUPPLY
Compressed Air: Between 2,5 to 5 Kg/Cm2
Oxygen: Between 2,5 to 5 Kg/Cm2
Maximum difference of pressure between both gases: Less than 1,5 Kg/Cm2.


NOTE

All equipmentes complies with IEC 60601/1 and IRAM 4220 requeriments.

 
 
   